Burst (EFT) 5/50 ns:
Pulse conforms to IEC/EN 61000-4-4
Burst frequency:
Adjustable from 1 kHz to 1 MHz up to 4.4 kV; 1 kHz to 250 kHz from 4.4 kV to 8.0 kV
Burst time:
1 to 360 sec.; 1 second resolution 1 to 240 min.; 1 minute resolution 1 to 24 hours; 1 hour resolution
Polarity:
Positive, negative, alternating
Pulse amplitude:
200V to 8.0kV, ±10%; 5 V resolution
Standards Met:
IEC 61000-4-4, IEC 61000-4-5, IEEE C62.41, UL 864, ANSI/IEEEC62.41, UL 1449
Manufacturer:
Thermo Keytek
Online Orders Ship Same Day

 

Availability:

In Stock

SKU:
ECAT

 

This product is available for shipping to North America and Western European countries.


Getting Started...

Check out our Google Reviews

Overview

ECAT Model E103 Control Center (Included with Modules)

Computer-driven control center and power units required for the operation of any Thermo Scientific ECAT test system.

Surge V&I Monitor Ports - For waveform monitoring with an external (user supplied) oscilloscope at 1kV/V and 200A/V;V & I signals supplied from optional AC coupler/decoupler or ECAT modules/options.

Input Voltage: 100V to 240VAC, 50/60 Hz

Typical Input Current: 3.5A @ 100V; 1.5A @ 240V

ECAT Model E510A Module

Plug-in module to produce combination wave specified by ANSI/IEEE C62.41 Cat. B and IEC 61000-4-5 to 10kV and 5kA.

Compatible Coupling/Decoupling Networks: E455x-kV, E4555, E4556

ECAT Model E506-4W Module

Plug-in module to produce the 2/10µs surges required by Telcordia GR-1089 CORE for up to five-wire (four terminal) testing.

Waveforms:

  • <2/>10µs, 50-800V, 100A/terminal with 800V open circuit voltage
  • <2/>10µs, 50-1500V, 100A/terminal with 1500V open-circuit voltage
  • <2/>10µs, 100-2500V, 500A/terminal with 2500V open-circuit voltage
  • <2/>10µs, 200-5000V, 500A/terminal with 5000V open-circuit voltage

Tolerances: All peak open-circuit voltages and short-circuit currents are -0%/+20%

Outputs: Front panel terminals for connection to T1, R1, T2, R2 and Ground

Surge: Repetition Rate 1 shot/16 seconds

ECAT Model E508 Module

Plug-in module to produce the 10/360µs surges required by Telcordia GR-1089 CORE.

Open-Circuit Voltage: 10/360µs, 50-1100V, -0/+15% in 1 volt steps. Tip and ring outputs independent and isolated to insure true, three-terminal simultaneous testing of up to 12 pair. Waveforms as defined by Telcordia GR-1089-CORE.

Short-Circuit Current: 100A/side -0/+15% at a voltage setting of 1.0kV

Front Time Tolerance: -30/+0% for voltage and current

Duration Tolerance: -0/+30%

Surge Repetition Rate: 1 shot/50 seconds

ECAT Model E508-12P Module

Plug-in module to produce the 10/360µs surges required by Telcordia GR-1089 CORE.

Open-Circuit Voltage: 10/360µs, 50-1100V, -0/+15% in 1 volt steps. Tip and ring outputs independent and isolated to insure true, three-terminal simultaneous testing of up to 12 pair. Waveforms as defined by Telcordia GR-1089-CORE.

Short-Circuit Current: 25A/side -0/+15% at a voltage setting of 1.0kV

Front Time Tolerance: -25%/+0 for voltage; -30%/+0 for current

Duration Tolerance: -0/+30%

Surge Repetition Rate: 1 shot/150 seconds

ECAT Model E411 Module

EFT/Noise Burst simulator for testing in accordance with IEC 61000-4-4 Edition 2 to 4.4kV.

Burst Polarity: Positive, negative, alternating

Burst Voltage: 200 V to 4.4 kV

Burst Voltage Resolution: 5 V

Coupling/Decoupling Network: E455X

ECAT Model E502B Module

Plug-in module to produce the telecommunications surge wave required by IEC 61000-4-5, FCC Part 68 and ITU Rec K.17, K.20, and K.21 (formerly CCITT).

Open-Circuit Voltage: 10/700µs and 0.5/700µs, 200V - 6.6kV ±10% in 1 volt steps.10/700µs waveform meets both IEC and FCC Part 68 9/720µs requirements. Tighter tolerances for front time and duration ensure compliance with both requirements.

Short-Ciruit Voltage: Open-circuit voltage ÷ 15 with 0 ohm effective source impedance; open-circuit voltage ÷ 40 with 25 ohm effective source impedance. Tolerance is -0/+10%

Front Time Tolerance:

  • Voltage: 7.0µs to 11.7µs
  • Current: 5.0µs ±30%

Duration

  • Voltage: 576µs to 840µs
  • Current: 320µs ±20%

Surge Repetition Rate: 1 shot/18 seconds

ECAT Model E503 Module

Plug-in module to produce the ring waves specified by ANSI/IEEE C62.41 Cat. A and B, and various UL standards, including UL 864.

Voltage Rise Time: 0.5µs ±30%

Ringing Frequency: 100kHz ±20%, 40% decay per peak

Open-Circuit Voltage: 200V - 6.6kV ± 10%

Short-Circuit Current: Selectable at 200A max. or 500A max., when the open-circuit voltage is set to 6.0kV. (Actual shortcircuit current at other voltage settings will be open-circuit voltage ÷ 30 when 200A is selected and open-circuit voltage ÷ 12 when 500A is selected.)

Surge Repetition Rate: 1 shot/9 seconds

Line Sync Accuracy: ±15º with optional coupler/decoupler

ECAT Model E000 Blank Module

Blank Plug-in Module. Required when only a single 1/2 width module is installed in a docking bay.

 

About the Keytek ECAT System:

The Thermo Fisher Keytek ECAT Simulator systems are a sesoned favorite being used for electromagnetic Surge and EFT testing for more than 2 decades. The controller has an easy-to-use Windows based application software for quick implementation of test routines and standards, with automoatic report generation. A multi-level system interlock architecture provides maximum safety to keep first time users safe.

This rental ECAT system is suitable for users looking to evaluate DUT in a precompliance environment for passing Surge and Burst (Electrical Fast Transient) immunity test standards such as IEC 61000-4-4, IEC 61000-4-5 and ANSI C62.41. The single phase, 15 amp coupling network is suitable for most commercial electronics and lower powered electronics.

Common Configuration for ECAT:

  • E103 Controller (fiber-interface for remote control)
  • E421 EFT/Noise Burst Simulator to 8 kV (IEC 61000-4-4)
  • E4554 Burst Coupling up to 8 kV
  • Input voltage: 100V to 240VAC
  • Input Current: 3.5A at 100 V; 1.5A at 240 V
  • User Manual
  • NIST Calibration Certificate

Power Line Coupling/Decoupling Networks

Surge Data Line Couplers

EFT Couplers

Waveform Verification and Calibration Equipment

Surge & EFT Test Setup

ISO 17025 ACCREDITED - ONSITE SCHEDULING

Calibrations & Repairs

ISO 17025
Scope of Accreditation

Videos

Key Considerations

Manufacturer:
Thermo Keytek
Burst (EFT) 5/50 ns:
Pulse conforms to IEC/EN 61000-4-4
Burst frequency:
Adjustable from 1 kHz to 1 MHz up to 4.4 kV; 1 kHz to 250 kHz from 4.4 kV to 8.0 kV
Burst time:
1 to 360 sec.; 1 second resolution 1 to 240 min.; 1 minute resolution 1 to 24 hours; 1 hour resolution
Polarity:
Positive, negative, alternating
Pulse amplitude:
200V to 8.0kV, ±10%; 5 V resolution
Standards Met:
IEC 61000-4-4, IEC 61000-4-5, IEEE C62.41, UL 864, ANSI/IEEEC62.41, UL 1449