AE Techron T1000 Overview
The T1000 Magnetic-Field Susceptibility Transformer was designed to exceed the AF Magnetic Field susceptibility test requirements of DO 160 Section 19 and Boeing D6-16050 Section 7.2. It may also be used as a 10-to-1 step-down transformer for output current of up to 100ARMS
The T1000 transformer provides up to 300A-m and is able to generate up to 100ARMS. The turns ratio provides a ten-to-one step down to the secondary winding.
The T1000 provides convenient input connectors via binding posts. Standard 0.75-inch spacing of binding posts allows use of standard plugs. Output is via Multi Contact 125A plugs.
A 12-amp circuit breaker protects the unit against over-current conditions. A rugged, impact-resistant case and robust design protects the transformer from accidental damage.
The T1000 transformer is for use by experienced staff.
[split]
- Up to 100 ARMS secondary current
- Provides up to 300 A-m
- Ten-to-one step-down
- Output via Multi Contact 125A plugs; mating connectors included
Controls and Connectors
Input Terminals: 4-way binding posts
Output Terminals: Multi Contact 125A plugs
Circuit Breaker: Primary input limited to 12A; push to reset
Primary: 200 Vp maximum input
Secondary: 20.0 Vp maximum output
Dielectric Test: 1200 VDC
Turns Ratio: Ten-to-one step down
Protection: Input power-protected via circuit breaker