Overview
The Teseq ITS 6006 consist of an integrated RF signal generator, RF switch and EUT monitoring interfaces. The unit is designed for various radiated EMC applications in the 80 MHz to 6 GHz frequency range. In addition to the generator, the system includes the AM and PM modulators necessary for EMC testing. The second generation, called ITS 6006B, includes 3 freely configurable pulse modulators (1 µs to 200 s) for radar pulse profiles as required e.g. by Ford FMC1278 or VW TL81000. The RF signal can be switched to one of four outputs, where up to four power amplifiers can be connected directly. Two RF switches are supplied for combining four amplifier into two antenna paths.
ITS 6006 interlock outputs ensure the activation of the power amplifier for only the selected signal path. The 3.5“ color display shows the generator parameters. LEDs indicate the state of further functions.
The EUT monitoring is provided by six digital, one analog and one optical input, as well as four digital outputs. Up to three power meters PMR 6006 can be connected directly to the ITS 6006. The ITS 6006 provides remote control through its network, RS232 and USB interfaces.
Parameter | Value |
---|---|
Frequency range | 80 MHz – 6 GHz |
Resolution | 1 Hz |
Reference frequency | 10 MHz |
Max. adjustment deviation | ±50 Hz |
Aging | 5 ppm |
Level range | -60 dBm to +10 dBm (-60 dBm to +8 dBm at f>3 GHz & ϑ>30°C) |
Resolution | 0.1 dB |
Accuracy | <1 dB, typ. <0.5 dB (+15 to +25°C) |