Overview
Radiating Loop 120 mm (acc. to MIL-STD-461E RS101) including 3m Cable, Calibration Sheet.
For immunity tests radiating loops are necessary to generate magnetic fields. Suitable loops are available. Measuring emissions require loop sensors which can also be ordered from our company.
SPECIFICATIONS | ||||
Mechanical Data | Loop Sensor MGA_LS040 | Radiating Loop MGA_RL120 | Loop Sensor MGA_LS133 | Loop Sensor / Radiating Loop MGA_RLS133 |
Diameter | 40 mm | 120 mm | 133 mm | 133 mm |
Body material | PVC | MDF | MDF | MDF |
Wire | 7-41 Litz wire | 2.0 mm copper wire | 7-41 Litz wire | 0.25 mm² Litz wire |
Number of turns | 51 | 20 | 36 3 | 6 |
Number of layers | 1 | 4 | 4 | 4 |
Shielding | Electrostatic | none | Electrostatic | Electrostatic |
Distance to EUT | 5 cm | 5 cm | 7 cm | 10 cm / 5 cm |
Connector at coil | Speakon | 4 mm MC socket | Speakon | Speakon |
Connector at cable | XLR | 4 mm MC plug | XLR | XLR/4 mm MC plug |
Coil factor (50 mm) | --- | 76,3 1/m | --- | 138,5 1/m |
Electrical Data | ||||
Correction factor | see calibration sheet (50 ? / 600 ? / 1M?) | --- | see calibration sheet (50 ? / 600 ? / 1M?) | see calibration sheet (50 ? / 600 ? / 1M?) |
DC resistance | ~ 4,5 ? | ~ 0,05 ? | ~ 10 ? | ~ 1,1 ? |
Inductance | ~ 130 µH | ~ 120 µH | ~ 340 µH | ~ 340 µH |
Resonant frequency | --- | > 1.8 MHz | --- | > 0.9 MHz |
Frequency range | 10 Hz - 1 MHz | DC - 500 kHz | 10 Hz - 1 MHz | DC / 10 Hz - 500 kHz |
Nominal current | --- | 15 A | --- | 5 A |
General Data | ||||
Connecting cable | Microphone cable | Litz wire 2 x 1.5mm² | Microphone cable | Microphone cable / Litz wire 2 x 1.5mm² |