Em Test CWS 500N1.4 RF Conducted Immunity Test System for IEC 61000-4-6

Test Capabilities:
RF Conducted Immunity
Standards Met:
IEC/EN 61000-4-6
Applications:
Radiated Immunity Debugging, CE Marking, Medical Devices, Industrial Machinery, Electronics & IT
CDN Test Levels (w/ built-in amplifier):
40Vrms
BCI Test Levels (w/ built-in amplifier):
8Vrms
Frequency Range (w/ built-in amplifier):
150 kHz - 230 MHz
Manufacturer:
EM Test
Online Orders Ship Same Day

 

SKU:
sku-251

 

This product is available for shipping to North America and Western European countries.


Getting Started...

Check out our Google Reviews

Overview

The CWS 500N1.4 is the most compact single box test equipment for testing conducted rf immunity per IEC 61000-4-6 Ed. 3 and Ed. 4 and related standards with a frequency range of 100 kHz to 300 MHz. Apart from the 1 kHz 80 % AM signal the generator also generates a 2 Hz 80 % AM signal to test medical appliances and a 1 Hz PM signal with 50 % duty cycle required to test safety equipment like fire alarms. Equipped with a 1 GHz current monitor the CWS 500N1.4 can be used up to 1 GHz by means of an external amplifier. EM TEST supplies a large range of CDNs, EM clamp and current injection clamps as well as the corresponding calibration accessories. Full compliant levelling can be run from the front panel of the CWS 500N1.4 storing the results in 5 memory spaces is available.

Operating Functions

Front Panel

EM Test CWS 500 Front View
1 "Test On"
By pressing the key "TEST ON", the test procedure can be started with the pre-selected parameters and amplifier is turned ON.
7 Escape
The ESC button returns back to the previous level in the menu.
2 Function Keys "F1 .. F7"
Parameters and functions displayed in the lowest line and functions displayed with ”F “, can be selected with the related function key.
8 LED Current Probe Input
At the BNC output the current pulse of the generator can be measured. Dividing ratio is 10:1 (5V at 50A pulse).
3 Display
All functions and parameters are displayed (8 lines with max. 40 characters).
9 LED RF Output
When the RF output is active this LED on the front panel is illuminated to indicate that a test signal is applied to the test object.
4 EXIT
The EXIT button resets the firmware to the main menu. This is only possible, if no test routine is running.
10 RF-Output
At this output the RF power is available. The 6dB-attenuator is connected via coaxial cable. For conducting tests together with CDN’s, EM clamps or current injection clamps, the generator must always be loaded with a 6dB-attenuator as per IEC 61000-4-6.
5 Cursor key
Parameters and functions can be changed during the test. The selection of these parameters is done with the cursor, moving it to the left or to the right. The increment value of the selected parameters in the lowest line can also be selected with the cursor key.
11 Current Probe Input
For tests with a coupling clamp where the EUT current has to be monitored, the current probe can be connected to this input.
6 Exit Knob (Inc / Dec)
This knob increments or decrements test parameters with a numeric value or selects parameters from a list.
ATTENTION: Do not connect the RF output directly to this input otherwise the CWS500 might damage.

 

Rear view

EM Test CWS 500N1.4 Rear View
1 Bridge AMP OUT – Coupler IN

7 IEEE Interface
2 Air output

8 USB Serial Interface
3 Bridge SG OUT – AMP IN 9 Remote Control Connector
Not used.

4 Safety Circuit 10 Fail 1 Detection "Stop"
The BNC input FAIL 1 is for failure detection on the EUT. If the input is set to ground (chassis), the
CWS 500N1.x generator stops the actual test routine and terminates the test. It is not possible to continue
the test routine.

A complete restart of the routine will be necessary. The message "FAIL 1" indicated on the LCD-Display as
well as in the icd.control software.
5 Power On Switch and Fuse 11 Fail 2 Detection "Pause"
The BNC input FAIL 2 is for failure detection on the EUT. If the input is set to ground (chassis), a failure event detects, and the test continues normally. After 10 FAIL 2 events the program will stop and the message "FAIL 2: 10" indicates on the LCD-Display as well as in the icd.control software.
6 Power selector 115V – 230V

 

Technical Data
Test Level CWS 500N1.3
Output Level 1 - 30Ve.m.f. after 6dB-attenuator
Output power 80W (nominal)
Output impedance 50 Ohm
max. VSWR 1:1.2 at all phase angles and at max. power (without destruction)
Harmonic distortion
(at max. power)
< 15dBc
Test frequencies CWS 500N1.3 CWS 500N1.4
Sinusoidal (CW) 10 kHz – 400 MHz 100 kHz – 300 MHz step: 1kHz
  In the menu QUICK START the step size can be selected by the user
Modulation
Modulation method Amplitude modulation
80% <±5% and 1kHz <±10% acc. to IEC 61000-4-6
80% <±5% and 2Hz, 1kHz acc. to IEC 60601-1-2
Pulse modulation 1Hz, 50% duty cycle acc. to EN 50130-4
General 80% at Frequencies 2Hz, 400Hz,1kHz
Other modulation via ICD software Settings in Vector mode
Signal Generator
Output level -63.5 dBm to 0 dBm
Step level 0.5 dB
Frequency range 9 kHz to 1 GHz
Output impedance 50 ohm
Calibration measurements
Cal data 5 internal stores for calibration data.
ICD Software for the upload and download of Cal-files.
LCD On-line display of the test level and the preselected frequencies.
Monitor input PM1000 RF monitor input for calibration procedures and current probes measurement. Input voltage for current monitoring V max = 1.0 V rms (13dBm)
Cal data for current probe (CP) 2 internal stores ; ( download with ICD software )
A download will reduce data to 10/ decade and frequency range 100kHz ... 300MHz
Timing
Dwell time general td = 0,3s - 9999.9s
pulse modulation td = 3s - 9999.9s
Rest time tr = 0 / 0,3s - 9999.9s
Output
Direct N
EUT control
BNC input FAIL 1 Fail 1; test will be stopped immediately
BNC input FAIL 2 Fail 2; failure is detected, test continues (max. 10 failures)
Test routines (integrated)
Quick Start Immediate start, all parameters adjustable during testing.
User test routines 1. Voltage sweep
2. Frequency sweep
3. Dwell time sweep
Standard test routines. 1. Level 1-3 acc. to IEC 61000-4-6
2. Automatic Level X - Level Y
Cal procedure Calibration of the complete test set-up. Internal storage of the Cal-files.
Service Service, Setup
Interfaces
Serial Interface USB
Parallel IEEE 488 GPIB interface Addresses 1 - 30 selectable
All interfaces are included as standard features.
Power meter PM 1000
Frequency range 9kHz – 1000 MHz
Input range Monitor -45 dBm ... +13 dBm
FWD power -50 dBm ... +13 dBm
REV power -50 dBm ... +13 dBm
Max. Input MONITOR 13dBm approx. 1.0 Vrms
Accuracy < ± 0.5 dB ( 0.09MHz...400MHz )
< ± 1.0 dB ( 0.01MHz...1000MHz )
Dual Directional Coupler
Frequency range 10kHz – 1000 MHz
Power 150 W max.
Insertion loss 0.6 dB max.
Mainline VSWR 1.1:1 max.
RF Amplifier ( build in ) CWS 500N1.3 CWS 500N1.4
Class of Operation Class A Class A
Frequency range 10 kHz .. 400 MHz 100 kHz .. 300 MHz
Output power nominal 80 W 80 W
Gain 48 dB nominal 48 dB nominal
Input power for rated output 0 dBm 0 dBm
Input power max. 10 dBm 10 dBm
Input / Output Impedance 50 Ω 50 Ω
Input VSWR 1.5: 1 max. 1.5: 1 max.
General Data
Dimensions 19"/3HE 19"/3HE
Weight 17.25 kg 14.95 kg
Power supply 110-230V/ max 50/60Hz
Input power max. 380W
Power factor cosΦ=0,98 at max output power acc. to IEC 555
Fuse 230V: 2x3,15AT
115V: 2x6.3AT
Cooling Active cooling, air ventilation
Environment conditions 10°C - 35°C 10°C - 40°C
Humidity Max. 85 %, non condensing

 

How to Order the EM Test CWS 500N1.4:

  • Click "Add to Cart" above
  • Register or Check Out as a Guest
  • Select Shipping Method (use "My Account" and input your account # and method)
  • Agree to Terms & Conditions
  • Payment Method Options: Credit Card or Submit A Purchase Order
  • Order Confirmation and Tracking Info Follows

Key Considerations

Manufacturer:
EM Test
Test Capabilities:
RF Conducted Immunity
Standards Met:
IEC/EN 61000-4-6
Applications:
Radiated Immunity Debugging, CE Marking, Medical Devices, Industrial Machinery, Electronics & IT
CDN Test Levels (w/ built-in amplifier):
40Vrms
BCI Test Levels (w/ built-in amplifier):
8Vrms
Frequency Range (w/ built-in amplifier):
150 kHz - 230 MHz

Connect with The EMC Shop

Our experienced staff can help you with applications, purchase and rental options.