APREL MiNi-HAC Hearing Aid Compatibility Automated Test System

Online Orders Ship Same Day

 

SKU:
sku-1592

 

This product is available for shipping to North America and Western European countries.


Getting Started...

Check out our Google Reviews

Overview

MiNi-HAC™ Hearing Aid Compatibility Automated Test System

Hearing Aid Compatibility (HAC) evaluation of wireless handsets ensures accessibility of these technologies to the hearing impaired.

The MiNi-HAC system can be used for the design and development of wireless handsets as well as for evaluation of hearing aids used by the hearing impaired to gain access to mobile wireless technologies.

APREL MiNi-HAC™ is the commercial compliance Hearing Aid Compatibility (HAC) bench measurement system, ideal as a primary compliance instrument for Hearing Aid Compatibility or as a design/development tool for electromagnetic near-field emission evaluation of antennas and wireless devices. The system is designed for developers, compliance experts, regulators, and researchers.

APREL is a pioneer in the area of Hearing Aid Compatibility, and the MiNi-HAC™ brings APREL’s almost three decades of experience into one package. APREL is ISO 17025 accredited for HAC standards including ANSI/IEEE C63.19.

MiNi-HAC™ is an affordable and easy-to-use system for complex HAC evaluations of wireless products (handsets, smartphones, etc.). It is an expandable system which can test devices operating up to 6 GHz and which will grow with technological developments.

This product is ideally suitable for use in a fast paced development area where assessments are needed for research & development, for pre-compliance and for final product compliance testing.

Basic system configuration

  • Software platform based on industry-recognized test methodologies and FCC-mandated test standards (ANSI/IEEE C63.19)
  • Software audio band testing module
  • Automated X,Y,Z precision probe movement system
  • Full graphic package for visualization and manipulation of measured data
  • E-Field HAC probe with 2 Standard frequency calibrations
  • H-Field HAC probe with 2 Standard frequency calibrations
  • Active Twin Axis T-Coil probe
  • 2 Standard frequency validation dipoles
  • Telephone Magnetic Field Simulator (TMFS) for reference check of the HAC T-Coil measurement setup
  • Helmholtz Coil (optional)
  • Device Positioner
  • Communications and Control Expander with integrated DAQ-PAQ and emergency stop mechanism
  • 5 axis laboratory-grade robot
  • Bench area of system under 6 square feet

Description Performs HAC testing for wireless communications devices (WD) in compliance with ANSI/IEEE C63.19 methodologies
Software User friendly GUI that allows for easy setup and data retrieval
Includes signal generator (no additional equipment needed)
Full FFT (Fast Furrier Transform) analysis
Multiple Octave Settings (user defined or standard)
Tone generation (comes with ITU P50)
Applications RF Emissions Test - Measurements of the near-field electric and magnetic fields emitted by a WD
RF E-Field emissions V/m
RF H-Field emissions A/m
AF T-Coil tests
ABM1 intensity and frequency response
ABM2 accurate evaluation
Typical Test Bands SMR 800, SMR 900, Cellular 850, GSM 850, 900, 1800, 1900
Custom bands available on request
Report Generation MS Word report generated according to FCC requirements
E-Field Probe E-Field Probe used for RF emissions testing of a wireless device
Typical sensitivity: 1.0 mV / (V/m)
Typical isotropy: 0.2 dB
H-Field Probe H-Field probe used for RF emissions testing of a wireless device
Typical sensitivity @ 835 MHz: 200 mv / (A/m)
Typical sensitivity @ 1880 MHz: 440 mv / (A/m)
Typical isotropy: 0.3 dB
Active Twin-Axis T-Coil Probe Converts magnetic fields of audio frequencies into AC voltage.
Built in preamplifier module which allows for extended dynamic range
Designed to sense axial and radial fields
Typical frequency range: 100 Hz to 20 kHz
Typical sensitivity: -60 dB(A/m) to 20 dB(A/m)
Validation Dipoles Typical performance is better than -10dB
Custom dipoles available on request
Magnetic Dipole TMFSTM Used to calibrate and validate automated methods used for HAC audio band magnetic system testing (ALSAS-10U HAC Upgrade SuiteTM)
Serves as a known source for audio band magnetic fields for system validation and calibration
Audio DAQ-PAQ and Integrated Power Amplifier All-in-one box
ADC (DAQ-PAQ) 16 Bit
Amplifier Range up to 100 dB input
Output @ 50 ? relative to 1 Khz -4 dBV
Communications via USB
Field Integration Local Co-Processor utilizing proprietary integration algorithms
LED Indication Emergency stop and DAQ-PAQ state
Number of Input Channels 4 in total: 3 dedicated and 1 spare for future upgrades
Communication Packet data via RS232 and USB
Ambient Noise 20 dB below intended measurement limit
Power Robot and controller supplied by 110 or 220 V standard (country specific) supply
Communications and control expander supplied by dedicated DC source (no battery required)

Connect with The EMC Shop

Our experienced staff can help you with applications, purchase and rental options.