Fischer/FCC F-10 Current Probe

Condition:
Used
Connector:
N-Type
Frequency Range:
10 Hz to 3 MHz
Applications:
Conducted Immunity
Standards Met:
IEC 61000-4-6
Transfer Impedance:
0.25 Ω (± 2 dB) from 2 kHz to 2 MHz
Max Primary Current:
350 Amps at DC to 400 Hz
Manufacturer:
FCC
Online Orders Ship Same Day
$395.00

 

SKU:
8161-RE
Availability:

In Stock


Getting Started...

Overview

The EMC Shop is an ISO 17025 accredited calibration company with the Fischer/FCC F-10 Current Probe in stock available for sale or rental. Repair and refurbishment services available as well. Contact us now for application support.

The Fischer/FCC model F-10 is used to measure currents on 50 Hz, 60 Hz and 400 Hz power lines. A power absorbing terminator is used to offer high sensitivity with broad bandwidth and high current handling capability. The F-10 current probe has a transfer impedance of 0.25 Ω (± 2 dB) from 2 kHz to 2 MHz. It has a usable frequency range overlapping the fundamental powerline frequencies. The F-10 can handle 350 amperes of primary power line current from DC to
400 Hz without saturation.

Videos

Key Considerations

Manufacturer:
FCC
Connector:
N-Type
Frequency Range:
10 Hz to 3 MHz
Applications:
Conducted Immunity
Standards Met:
IEC 61000-4-6
Transfer Impedance:
0.25 Ω (± 2 dB) from 2 kHz to 2 MHz
Max Primary Current:
350 Amps at DC to 400 Hz

Connect with The EMC Shop

Our experienced staff can help you with applications, purchase and rental options.