Overview

The 3ctest S6/S10/S20 Series is based on the third generation of intelligent multi-function comprehensive simulator. Set the test requirements on the parameter page and all the test procedure can be done automatically with the advantages of integrated system, intelligent control, easy-operation and time-saving. It fully meet the requirements of IEC 61643-1, GB18802.1, UL1449, YD-T1235[1].2 and TB/T 3074 surge standards for MOV, TVS and GDT surge protection device test and for mains surge protection test or scientific research.

As per

  • IEC 61643-1
  • GB18802.1
  • UL1449
  • YD-T1235[1].2
  • TB/T 3074
Technical Parameters
Model S6 S10 S20
Output waveform 8/20μs?other waveform please see attached table?
Front time T1 8μs±10%
Wave tail time T2 20μs±10%
Current peak Ip 100A?1200A?6kA?10kA 10kA?15kA?20kA 20kA?30kA?40kA
Current output range 10%-100% 1kA?100% 1kA?100%
Equivalent output impedance 50Ω?5Ω?1Ω?0.5Ω 1Ω?0.75Ω?0.5Ω 1Ω?0.75Ω?0.5Ω
Charging voltage Up to 6kV Up to 10kV Up to 20kV
Internal grade energy I grade III grade, automatic switch III grade, manual switch
Voltage polarity +/- (automatic switch)
Impulse interval 10s?999s
Number of continuous impulse 1?99
Charging voltage instability Less than 1%
Trigger mode Automatic or manual, single trigger available
Trigger malfunction rate Less than 1%
Trigger range 100% rated charging voltage
Measurement display resolution Better than 3%
Insulation strength ≤500V RMS?common port COMMON to ground reference PE?
General Data
Power supply AC220V±10%?50/60Hz?10A
Temperature -10-+40°
Humidity 35%-85%RH?no condensation?
Ground resistance No more than 0.5Ω
Dimension 19 inch 6U 19 inch 22U 19 inch 35U
Weight Approx. 35kg Approx. 120kg Approx. 200kg