Overview
Surge pulses occur due to direct or indirect lighting strokes to an external circuit. This leads to current or electromagnetic fields causing high voltage or current transients. Another source for surge pulsed is switching transients originating from switching disturbances and system faults. Nearly every electrical and electronic device may suffer from such surge pulses. CWS series is a high-end test instrument which fully complies with CE and CCC relative standards.with external CDN and optional accessories, it can conduct for three-phase DUT and also pulsed magnetic field immunity test. For CWS X00G/SPN38XXT, X stands for voltage in CWS X00G and X stands for network load current capacity in SPN38XXT.
General Data | ||
Display screen | 5.7inch TFTcolor touch screen | |
Power supply | AC 110V/220V ±10%, 50/60Hz | |
Fuse | 6 A | |
Max. Power consumption | 250 W | |
Memory space | Infinite ?PC? | |
Communication mode | Ethernet, RJ 45 | |
External control | D-sub 25p | |
External trigger input | BNC,5V TTL | |
CROtrigger output | BNC,5V TTL | |
Input | BNC,5V TTL | |
Synchronization | 45-65Hz,20-500Vbetween dotted terminal & undotted terminal 0-500Vamong dotted terminal, undotted terminal and case | |
Warning LED | 0-220VAC 10A/0-48VDC 3A Depending on external signal lamp | |
Safety circuit | Test only can be started when it is closed | |
Failure test | Display on LCD and interrupt the work at failure case | |
Working indicator | LED, LCDfront panel display | |
Grounding connection | Flat ground line | |
Output mode | Standard 1m coaxial cable | |
Size | 19inch/6Ucabinet | |
Weight | 35kg | |
Temperature | 15–35° | |
Humidity | 45%–75% | |
Air pressure | 86kPa–106kPa | |
Surge test as per IEC 61000-4-5 | ||
Standard | IEC / EN 61000-4-5?GB / T 17626.5 | |
Test voltage | 0.2kV–6.0kV/8.0kV?±10% | |
Voltage waveform | Front time: 1.2us±30%, Time to half value: 50us±20% | |
Test current | 0.1 kA–3.0kA/4.0kA?±10% | |
Current waveform | Front time: 8us±20%, Time to half value: 20us±20% | |
Source output impedance | 2 ohm | |
Output polarity | +/-/alternative | |
Coupling capacitance | 9uF, 18uF | |
Coupling Resistance | 10 ohm, 0 ohmcoupling resistance IEC standard mode or customized mode | |
Pulse cycle | 5s-99s?shortest depends on test voltage? | |
Number of test | 1–999 | |
Trigger mode | Automatic, manual and external | |
Synchronization | 0°-360°, 1°step or random | |
Internal CDN Load Capacity | ||
Standard module | ||
Model | SPN 38XXT | External |
Coupling capacitance | 9uF / 18uF | |
Decoupling inductance | 1.5mH | |
Common mode resistance | 10Ω | |
Superimposed mode | Automatic programmable mode | |
Network capacity | Max: AC 380V phase current 16A/32A | |
Basic equipments | ||
CWS X00G/SPN38XXT simulator, three-core power cord, short connection plug, test cable, ground wire, fuse, coaxial cable, factory test report, warranty report and user manual. |