Test Capabilities:
RF Conducted Immunity
Standards Met:
IEC/EN 61000-4-6
Applications:
CE Marking, Medical Devices, Industrial Machinery, Electronics & IT
CDN Test Levels (w/ built-in amplifier):
40Vrms
BCI Test Levels (w/ built-in amplifier):
8Vrms
Frequency Range (w/ built-in amplifier):
150 kHz - 230 MHz
Manufacturer:
Teseq
Online Orders Ship Same Day

 

Teseq NSG 4070C-80 Downloads

Datasheet

User Manual

Quick Start Guide

 

Availability:

Typically In Stock

SKU:
sku-809

 

This product is available for shipping to North America and Western European countries.


Getting Started...

Check out our Google Reviews

Teseq NSG 4070C-80 Overview

Description of Teseq NSG 4070C-80:

The Teseq NSG 4070 is a multi-functional device for carrying out EMC immunity tests to accompany development and conformity testing in accordance to EN/IEC 61000-4-6, Namur and several automotive BCI standards. Anyone who spends a considerable amount of time on test level setting, connecting EUT monitoring or writing test reports can now carry out immunity testing in a much more efficient manner with the 4th generation of NSG 4070. Its modular set-up using internal or external amplifiers enables a large variety of applications. The powerful and easy to use firmware makes the NSG 4070 independent from an external PC and control software, however it can also be remote controlled for system operation. A state-of-the-art data transfer of test and measurement data for documentation is provided by USB stick to be plugged into the front panel.

[split]

Generator

  • RF
    • Frequency range: 4 kHz – 1 GHz
    • Resolution: 1 Hz
    • Reference frequency: 10 MHz
  • RF Level
    • Level range: -60 dBm to +10 dBm
    • Resolution: 0.1 dB
  • Amplitude modulation
    • Modulation depth: 0 – 100%
    • Modulation frequency range: 1 Hz – 50 kHz
    • Frequency resolution: 1 Hz
  • Pulse modulation
    • Rise / fall time (10% / 90%): < 1 μs
    • Modulation frequency range: 0.1 Hz – 1 MHz
    • Frequency resolution: 0.01 Hz
    • Duty cycle: 0.1% to 100%
  • External modulation
    • Delay time: < 1 μs / 180°
    • Period: min. 20 μs
    • Pulse width: min. 10 μs

Power Meter

  • Frequency range: 4 kHz to 1 GHz
  • Linear measurement range
    • channel 1: -35 dBm to +27 dBm (NSG 4070C-60:-40 dBm to +27 dBm)
    • channel 2,3: -45 dBm to +20 dBm
  • Max. input/no damageNoise level: >5 dB below the measurement range
    • channel 1-3: +28 dBm
  • Input return loss: >20 dB (below 500 MHz), >17 dB (500 MHz to 1 GHz)
  • Connector: BNC socket, 50 ?
  • Accuracy 10 to 30°C: <0.5 dB, typ. <0.3 dB

Power Amplifier

  • Nominal output power: 80 Watts
  • Frequency range: 150 kHz to 230 MHz
  • Input impedance: 50 ?
  • Output impedance: 50 ?
  • Input return loss: min. 10 dB
  • Output return loss: nominal min. 9.5 dB, 0 dB without damage
  • Gain: min. 50 dB
  • Gain flatness: max. +/- 3 dB
  • Saturated output power:
    • < 400 MHz 46.5 dBm
    • > 400 MHz 45.4 dBm
  • Linear output power:Max. input power without damage: max. +10 dBm
    • < 400 MHz 45.4 dBm
    • > 400 MHz 43 dBm
  • Harmonic distortion at linear output power: typ. < -20 dBc

 

Features & Specifications of Teseq NSG 4070C-45:

teseq_nsg_4070c_35_test_system_for_conducted_radiated_immunity_4_khz_to_1_ghz_signal_generator_35_watt_amplifier_1

teseq_nsg_4070c_80_integrated_signal_generator_4_khz_to_1_ghz_80_w_integrated_class_a_power_amplifier_9

Teseq NSG 4070C-80 Key Considerations

Manufacturer:
Teseq
Test Capabilities:
RF Conducted Immunity
Standards Met:
IEC/EN 61000-4-6
Applications:
CE Marking, Medical Devices, Industrial Machinery, Electronics & IT
CDN Test Levels (w/ built-in amplifier):
40Vrms
BCI Test Levels (w/ built-in amplifier):
8Vrms
Frequency Range (w/ built-in amplifier):
150 kHz - 230 MHz
Videos:
Calibration Example, Remote Operation Setup

Connect with The EMC Shop

Our experienced staff can help you with applications, purchase and rental options.