Overview
The design and performance of coupling/decoupling network is specified in IEC/EN 61000-4-6. It also illustrates the cables of each CDN and transmission signal.
CDN S1 is designed for single-core coaxial shielded line as per IEC/EN 61000-4-6.
As per
- IEC 61000-4-6
- EN 61000-4-6
- GB/T 17626.6
| Technical Parameters | |
| Model | CDN S1 |
| Standard | IEC/EN 61000-4-6 |
| Frequency range | 150KHz-230MHz |
| Max. AC voltage (line to ground) | 100V |
| Max. DC voltage (line to ground) | 150V |
| Max. current | 1A |
| RF input port (BNC) | <20V |
| EUT port | BNC |
| AE port | BNC |
| Common mode impedance (EUT) | |
| 150KHz-26MHz | 150Ω±20Ω |
| 26MHz-80MHz | 150Ω +60Ω /-45Ω |
| Differential voltage factor (RF input –EUT port) | |
| 150KHz-80MHz | 9.5dB±1dB |
| 80MHz-230MHz | 9.5dB±2dB |
| Common mode interference (RF port / AE) | |
| 150 kHz | >60 dB |
| 1.5 MHz | >60 dB |
| 30 MHz | >60dB |
| 80 MHz | >50 dB |
| 230 MHz | >40 dB |
| Optional Accessories | |
| CAL 100F | 150Ω to 50Ω calibration unit |
| CAL 150F | 150Ω calibration unit |
| CDN ADP BNC | Calibration adapter |
| General Data | |
| Weight | 2.0kg |
| Dimension | L×W×H 21×13×10cm |
| Enclosure | Aluminum |
| Temperature | 15°C - 35°C (operating condition) |
| Humidity | 45% - 75% (operating condition) |
| Package | Carton |