Overview
The calibration kit CAS ISN is made for function tests and remeasurements on impedance stabilization
network (ISN) for unshielded balanced pairs. Measuring parameters are common mode (CM) impedance,
phase angle, voltage division factor, decoupling attenuation (isolation) and longitudinal conversion loss
(LCL). Impedance stabilization networks (ISN) are defined in CISPR 22 and CISPR 16-1-2. The CISPR 16-1-2
gives additional requirements and provides examples and measurements for the networks. The ITU-T
recommendations G.117 and O.9 offers the background knowledge for measurements on symmetrical
telecommunication lines.
Features & Specifications of Teseq CAS ISN extension for T2:
- Frequency range: 100 kHz to 100 MHz
- Impedance for the network analyzer ports: 50 Ω
- Connector: BNC-type female
- Impedance for the balanced port: 100 Ω
- Connectors for the balanced ports and ground: 1 mm banana
- Inherent balance: as given in the diagram
- Return loss (150 kHz to 30 MHz): >20 dB
- Dimension (LxWxH): 87 mm x 29 mm x 27 mm
- Weight: approx. 95 g