Overview
EM Test TSS 500N10 Application:
Telecommunication networks are exposed to lightning events. Therefore telecommunication equipment being connected to the outside world need to have appropriate protection to show an acceptable immunity to surge transients in order not to fail in case of lightning events. Telecom Surge Simulators of the TSS 500N series are used to proof the immunity of telecommunications equipment.
The TSS 500N10 is used to perform tests as per EN/IEC 61000-4-5 and related standards and complies with the requirements of ITU-T and FCC 97-270 (part 68) for Surge B pulse. It also generates the Surge voltage pulse 1.2/50 us as per EN/IEC 60950
The TSS 500N10 is used to perform tests as per EN/IEC 61000-4-5 and related standards and complies with the requirements of ITU-T and FCC 97-270 (part 68) for Surge B pulse. It also generates the Surge voltage pulse 1.2/50 us as per EN/IEC 60950
EM Test TSS 500N10 Standards Met:
- EN 60950-1
- EN 61000-4-5
- FCC 97-270 (part 68)
- IEC 60065
- IEC 60950-1
- IEC 61000-4-5
- ITU-T K.17
- ITU-T K.20
- ITU-T K.21
- ITU-T K.28
- ITU-T K.45
EM Test TSS 500N10 Benefits:
The TSS 500N10 is a compact tester to generate the Telecom Surge pulses 10/700 us and 9/720 us as well as the 1.2/50us Voltage Surge pulse.The TSS 500N10 is not just the pulse generator but it comes with the built-in coupling networks for 2-wire and 4-wire test applications. The outputs can directly be connected to the Telecom ports of the DUT. For testing Telecom lines EM TEST offers the appropriate coupling networks for 4-wire or 8-wire lines.
Pre-programmed Standard Test routines allow best user convenience. Still the TSS 500N10 offers the Quick Start test routine where parameters can be changed on-line during the test to evaluate the susceptibility level of an individual DUT.
AS PER ITU AND ETS RECOMMENDATIONS | |
Pulse 1.2/50 μs | |
Voltage (o.c.) | 500 V - 10,000 V ± 10% |
Rise time*) | 1 us ± 30% |
Pulse duration*) 5 | 50 us ± 20% |
Energy storage capacitor | 1 μF |
Pulse 10/700 μs | |
Voltage (o.c.) | 500 V - 10,000 V ± 10% |
Rise time*) | 6.5μs ± 30% |
Pulse duration*) | 700 μs ± 20% |
Energy storage capacitor | 20 μF |
Polarity | Positive, negative or alternating |
Counter | 1 - 30,000 or endless |
*) definition of waveform parameters as per IEC 60469-1. As per IEC 61000-4-5 this is considered to be equal to the waveform parameter definition as per IEC 60060-1 for the 1.2/50μs pulse and CCITT for the 10/700μs pulse. | |
AS PER FCC PART 68, PULSE B | |
Voltage (o.c.) | 500 V - 10,000 V ±10% |
Front Time | 9 μs ± 30% |
Decay time | 720 μs ± 20% |
Current (s.c.) | 720 us ± 20% |
Decay time | 320 μs ± 20% |
Energy storage capacitor | 20 μF |
Polarity | Positive, negative or alternating |
Counter | 1 - 30,000 or endless |
AS PER IEC 61000-4-5 | |
Pulse 10/700μs | |
Voltage (o.c.) | 500 V - 10,000 V ±10% |
Rise time*) | 10 us ± 30% |
Pulse duration*) | 700 μs ± 20% |
Current (s.c.) | 12.5 - 250 A |
Rise time*) | 5 μs ± 20% |
Pulse duration*) | 320 μs ± 20% |
Energy storage capacitor | 20 μF |