EN/IEC 61000-4-6

(Showing 12 of 24)

Overview

Testing to IEC 61000-4-6?

The EMC Shop stocks and rents this test equipment, allowing customers to preliminary evaluate their products before attempting certification. The IEC/EN 61000-4-6 test standard should be purchased, read and followed to develop a suitable test plan and effectively evaluate your product.

IEC 61000-4-6 is a standard for testing electromagnetic disturbances along data and power lines that run to and from your product (DUT = Device Under Test). While intimidating to decipher the 89 pages of edition 5 of EN/IEC 61000-4-6, it's testing techniques are simple and will help flush out other problem areas that arise in EMC testing. Most of our custommers usually call in, rent equipment and contact us with the equipment in hand. Dialing in the test setup, sequence or anything else is usually quickly resolved to get testing underway.

IEC 61000-4-6 Procedure, Setup & Testing

    • In lab or in place?
    • Determine test levels and coupling devices
    • Performance Criteria
    • Description of the EUT exercising method
    • Setup equipment and run calibration
    • Identify nearby and external contributors (if any)
    • Run test and identify problem areas
    • The recovery time of the EUT if necessary
    • Troubleshoot accordingly

Most of our customers will simply place the test generator on a bench, ground everything, and be off and rolling. There are diagrams avaialble that specify spacing, insulation, placement of the generator, etc... We have found that in most cases, these aren't detrimental in precompliance measurements when following IEC 61000-4-6. The test standard is long and detailed, covering different test setups, injection methods, procedures and more. The EMC Shop can support this effort with information on your DUT and surrounding auxilary equipment.

Selecting CDN's

Selecting CDN's According to IEC 61000-4-6

Test Equipment for IEC 61000-4-6

  • RF Conducted Immunity Test Generator
  • CDNs, EM Clamp or Current Clamp
  • Decoupling Device
  • Monitor Probe
  • 6dB Attenuator
  • 50 Ohm Terminations
  • Insulation Material
  • Grounding Plane
  • Calibration Kit (Optional)
  • RF Shield Room (Optional)

Calibration Setup

Calibration Setup Diagram acc. IEC 61000-4-6

Diagram - IEC 61000-4-6 CDN Test Setup

Test Setup

Test Setup Diagram acc. IEC 61000-4-6

Diagram - IEC 61000-4-6 Testing of Large Products

Video - Calibration Example (BCI)

Popular Systems for IEC 61000-4-6 Testing

Get Started

The EMC Shop supports all IEC 61000-4-6 testing equipment and setup.

Show 12 24

EN/IEC 61000-4-6 Test Equipment, Systems & Solutions

  • Monthly Rental Teseq (Schaffner) CDN T002 Coupling Network for Unshielded Balanced Pairs

    SKU: sku-318

    Application/Test Type: EMC testing for unscreened data communication/telecom lines (Power over Ethernet, 1100BaseT4, 100Base VG-AnyLan, 1000BaseT, Token ring, ISDN, 10BaseT, 100BaseTX, ATM, FDDI, TP-PMD
    Current Rating: 1.5 Amps (3/5 Amp models available)
    Frequency Range: 150 kHz to 230 MHz (limited to 80 MHz on 3/5 Amp models)
    Mains & EUT Connections: Banana, XLR, RJ45, RJ11
    RF Connector: BNC, 50 Ω
    RF Input Voltage: 30 Volts Max
    Standards Met: IEC 61000-4-6
    Voltage Rating: AC max. voltage (line to ground): 63 Volts DC max. voltage (line to ground): 100 Volts Test voltage: 200 VDC, 2 s
    $99.00
  • EMIC-20 EM Injection Clamp for IEC/EN 61000-4-6 Immunity Test

    SKU: 7851

    Standards: EN/IEC 61000-4-6
    Application/Test Type: Injects the disturbance signal through a combination of inductive and capacitive coupling
    Clamp factor (CF): -.5 dB at 100 kHz to -4 dB at 1 GHz (Min -7 dB at 700 MHz
    Connector: N-type female
    Dimensions: 632 mm x 120 mm x 80 mm
    Frequency Range: 10 kHz to 1000 MHz
    Max. Cable Diameter: 20 mm
    Maximum Continuous Power: 0.15 to 100 MHz: 100 W for 15 min; 100 to 230 MHz: 100 W for 5 min; 230 to 1000 MHz: 50 W for 3 min
    $595.00
    Typically In Stock
  • EMIC-35 Current Injection Clamp for RF Conducted Immunity Testing

    SKU: sku-2132

    Weight: approx. 14 kg
    Application/Test Type: Injects the disturbance signal through a combination of inductive and capacitive coupling
    Clamp factor (CF): -.6 dB at 100 kHz to -2.2 dB at 1 GHz (Min -4 dB at 700 MHz
    Connector: N-type female
    Dimensions: 666 mm x 135 mm x 120 mm
    Frequency Range: 10 kHz to 1000 MHz
    Impedance: 50 Ohms
    Max. Cable Diameter: 35 mm
    Maximum Continuous Power: 0.15 to 100 MHz: 100 W for 15 min; 100 to 230 MHz: 100 W for 5 min; 230 to 1000 MHz: 50 W for 3 min
    Standards Met: IEC/EN 61000-4-6; CISPR 24; EN 61000-6-1; EN 61000-6-2; IEC 60533; IEC 60601-1-2; IEC 61326
  • Teseq CDN S751A 75 Ohm BNC RF Coupler for IEC 61000-4-6 Testing

    SKU: sku-851

    Weight: 1.5 kg (3.3 lbs)
    Common Mode Impedance: 150 kHz to 26 MHz: 150 Ω ±20 Ω; 26 MHz to 230 MHz: 150 Ω +60 Ω / -45 Ω
    Current Rating: 1000 mA maximum
    Decoupling of Common Mode Disturbance: 150 kHz: Over 60 dB; 1.5 MHz: Over 60 dB; 30 MHz: Over 50 dB; 230 MHz: Over 40 dB (refer to data sheet for some exceptions)
    Dimensions: 4 x 4 x 9.5 inches (10 cm x 10 cm x 24 cm); (CDN HDMI and CDN S752A: 28.8 cm 11 inches)
    Frequency Range: 150 kHz to 230 MHz
    Mains & EUT Connections: BNC 75 Ω
    RF Connector: BNC 75 Ω
    RF Input Voltage: 20 Volts Max
    Standards Met: IEC 61000-4-6
    Voltage Division Factor: 10 dB ±1 dB to 80 MHz, 10 dB +3, -2 dB from 80 MHz to 230 MHz
    Voltage Rating: AC max. voltage (line to ground): 250V; DC max. voltage (line to ground): 400 V; Test voltage: 200 V - 750 V DC, 2 sec (refer to data sheet for exact model specifications)
    $1,245.00
  • AH Systems BCP-512 Probe to Measure Conducted RF Currents, 1 MHz - 1 GHz

    SKU: sku-255

    Weight: 0.5 pounds (227 grams)
    Aperture: 1.2" (32 mm)
    Application/Test Type: Conducted and Radiated Susceptibility Testing
    Current Rating: 200 Amps Maximum
    Dimensions: 2.8" (71 mm) diameter
    Frequency Range: 1 MHz to 1 GHz
    RF Connector: N-Type (female)
    Standards Met: IEC 61000-4-6; IEC 60601-1-2, Mil-STD-461 CS 114
    Transfer Impedance: 0 to 24 dBΩ
  • Com-Power CLCI-100 Bulk Current Injection Probe for 10 KHz to 100 MHz

    SKU: sku-384

    Weight: 4.5 lbs. (2.04 Kg)
    Aperture: 1.575” (40 mm)
    Application/Test Type: Conducted and Radiated Susceptibility Testing
    Dimensions: Window Diameter: 1.575” (40 mm) Outside Diameter: 5.75” (146 mm) Width: 2.75” (70 mm)
    Frequency Range: 10 KHz to 100 MHz
    RF Connector: N-Type (female)
    Standards Met: EC 61000-4-6 and CISPR 16-1-2